OSTEOARTHRITIS
Despite initial migration, Furlong Active stem comparable to Furlong HAC after 5 years .

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2015;3(6):29 Acta Orthop. 2014 Dec;85(6):556-61.
Auteurs contributeurs

E Weber M Sundberg G Flivik

50 patients with primary osteoarthritis undergoing total hip arthroplasty were randomized to receive either the Furlong Active femoral stem or the traditional Furlong HAC femoral stem. The purpose of this study was to evaluate the 3-D migratory patterns of the two stems using radiostereometry. Results after 3 months demonstrated significantly greater subsidence of the Furlong Active stem compared to the Furlong HAC stem, though subsidence of each stem plateaued after 3 months and displayed minor changes through to 5 years. Both groups displayed initial retroversion of the implant at 3 months, though minimal differences through to 5 years thereafter. Clinical outcome significantly improved in both groups over 5 years, with no significant differences detected between groups at 1, 2, or 5 year follow-up.


Détails du financement de la publication +
Financement:
Industry funded
Sponsor:
JRI Ltd.
Conflits:
None disclosed

Risque de partialité

6,5/10

Critères de déclaration

14/20

Indice de fragilité

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

Oui = 1

Incertain = 0,5

Non pertinent = 0

Non = 0

L'évaluation des critères de rapport permet d'évaluer la transparence avec laquelle les auteurs rapportent les caractéristiques méthodologiques et les caractéristiques de l'essai dans la publication. L'évaluation est divisée en cinq catégories qui sont présentées ci-dessous.

1/4

Randomization

2/4

Outcome Measurements

4/4

Inclusion / Exclusion

4/4

Therapy Description

3/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

L'indice de fragilité est un outil qui aide à l'interprétation des résultats significatifs, en fournissant une mesure de la force d'un résultat. L'indice de fragilité représente le nombre d'événements consécutifs qui doivent être ajoutés à un résultat dichotomique pour que le résultat ne soit plus significatif. Un petit nombre représente un résultat plus faible et un grand nombre un résultat plus fort.

Pourquoi cette étude était-elle nécessaire maintenant ?

The Furlong HAC stem is a well-proven femoral implant that has recently undergone modifications to the lateral fin to prevent fracture of the greater trochanter during insertion. The new stem - the Furlong Active stem - is a shorter cone-shape, double-tapered, and has more rounded medial and lateral edges. Using radiostereometry, this study aimed to compare the migratory patterns of the Furlong HAC and Furlong Active stem as a surrogate for osseointegration and long-term survival of the arthroplasty.

Quelle était la principale question de recherche ?

Do design modifications to the Furlong HAC stem to give the Furlong Active stem affect the migratory behaviour, osseointegration and long-term result of total hip arthroplasty up to 5 years after surgery?

Caractéristiques de l'étude +
Population:
50 patients (mean age: 63 years, mean BMI: 27, 27M/23F) with primary osteoarthritis scheduled for total hip arthroplasty. Patients who had unstable femoral anatomy as determined by radiographs and who were Charnley category C patients were excluded from the study. 45 patients (23M/20F) completed the study to completion at the 5 year endpoint. Surgery for both groups was performed using a posterolateral approach and 8-10 tantalum bone markers (diameter=0.8 mm). Apart from the femoral head used as a marker, all prostheses were pre-marked with tantalum markers at the tip, shoulder and medial collar of the stem.
Intervention:
Furlong Active group: Participants received the Furlong Active stem. Material was titanium alloy with a hydroxyapatite coating for uncemented implantation (n=25; 9M/16F) (Mean age: 62 years, range: 50-77 years)
Comparaison:
Furlong HAC group: Participants received the Furlong HAC stem. Material was titanium alloy with a hydroxyapatite coating for uncemented implantation (n=25; 18M/7F) (Mean age: 63 years, range: 49-75 years)
Résultats:
Radiostereometry (RSA) and conventional radiographic examinations were performed to analyze motion of the stem in relation to the tantalum markers implanted in the femur. Stem subsidence (Y-translation) and retroversion (Y-rotation) were primary effect variables. Self-administered Hip Disability and Osteoarthritis Outcome Score (HOOS) was also assessed.
Méthodes:
RCT: Single-centre
Durée de l'intervention:
Radiostereometry (RSA) and conventional radiographic examinations were conducted on the first postoperative day, before weight-bearing and then at 3 months, 1,2 and 5 years (+/-5% time tolerance) postoperatively. Routine clinical follow-up occurred 3 months and 1, 2 and 5 months postoperatively. Hip Disability and Osteoarthritis Outcome Score was evaluated prior to surgery and at 1, 2 and 5 years.
Quels sont les résultats importants ?
  • Significantly greater subsidence of the implant was noted in the Furlong Active group (0.99mm distally) compared to the Furlong HAC group (0.31mm distally) (p=0.0.46). Subsequent subsidence from 3 months to 5 years was not statistically significant in the Furlong Active (p=0.5) or HAC (p=0.06) stem.
  • There was one case of extreme subsidence reported in the Furlong Active group (6.5mm after 3 months) in a patient with BMI 41. Exclusion of this outlier case from the main analysis resulted in a non-significant difference in subsidence between groups (p>0.05).
  • Mean retroversion of the femoral stem was measured at 1.18deg in the Furlong Active group and 0.80deg in the Furlong HAC group after 3 months; the difference between groups was not statistically significant. (p=0.46). From 3 month to 5 years, no significant changes in mean retroversion were noted in either the Furlong Active (p=0.1) or HAC group (p=0.1).
  • Two patients, one in the Furlong HAC group (4.2 degrees) and one in the Furlong Active group (5.1 degree), demonstrated significant retroversion of the stem within the first 3 months after surgery. Between 3 months and 5 years, no further retroversion was noted in either patient.
  • No stem in either group was judged to be loose over the 5 year follow-up period.
  • HOOS scores improved significantly within both groups, without significant difference between groups at 1, 2, or 5 years after surgery.
  • Minor X- and Z-plane translation and rotation of the stem were noted in both groups, and differences were small between groups.
De quoi dois-je me souvenir en priorité ?

In total hip arthroplasty, subsidence of the uncemented Furlong Active femoral stem was significantly greater than the Furlong HAC stem in the first 3 months after surgery, although subsidence in both groups thereafter plateaued, with minimal subsidence occurring in either group from 3 months to 5 years. Both stems displayed initial retroversion of the femoral implant in the first 3 months, but minor changes thereafter to 5 years. No stem in either group was considered loose. Clinical outcome was similar between the Furlong Active and HAC stem throughout the 5-year follow-up.

Comment cela affectera-t-il les soins prodigués à mes patients ?

Though subsidence of the Furlong Active stem was greater than that of the Furlong HAC stem, early migration did not appear to negatively impact 5-year outcome. Namely, osseointegration of the implant was observed to occur after 3 months, with minimal subsidence migration of either implant from 3 month to 5 years, and no loose components observed. Additionally, retroversion did not appear to substantial differ between stem designs. These findings suggest that similar results may be attained with the new Furlong Active stem relative the standard Furlong HAC stem, and that the new design modifications do not negatively affect integration of the component or clinical results. Nevertheless, additional research in a larger cohort of trials should be considered to confirm the the findings of the current study.

AVIS DE NON-RESPONSABILITÉ

Le contenu de cette page est fourni à titre d'information uniquement et n'est pas destiné à remplacer un avis médical, un diagnostic ou un traitement professionnel. Si vous avez besoin d'un traitement médical, demandez toujours l'avis de votre médecin ou rendez-vous au service des urgences le plus proche. Les opinions, croyances et points de vue exprimés par les individus sur le contenu de cette page ne reflètent pas les opinions, croyances et points de vue d'OrthoEvidence.

0 de 4 articles mensuels GRATUITS débloqués
Vous avez atteint votre limite de 4 vues d'articles gratuits ce mois-ci

Accédez à OrthoEvidence pour seulement 1,99 $ par semaine.

Restez informé des dernières données. Annulez à tout moment.
  • Évaluations critiques des derniers essais contrôlés randomisés à fort impact et des revues systématiques en orthopédie.
  • Accès au contenu des podcasts OrthoEvidence, y compris les collaborations avec le Journal of Bone and Joint Surgery, les entretiens avec des chirurgiens de renommée internationale et les tables rondes sur l'actualité et les sujets orthopédiques.
  • Abonnement à The Pulse, une lettre d'information bihebdomadaire fondée sur des données probantes, conçue pour vous aider à prendre de meilleures décisions cliniques.
Upgrade
Close Dialog
Bienvenue à nouveau !
Vous avez oublié votre mot de passe ?
Commencez votre essai GRATUIT dès aujourd'hui !

Votre compte sera affilié à
et inclut un accès gratuit à OrthoEvidence.


OU
Vous avez oublié votre mot de passe ?

OU
Veuillez vérifier votre adresse électronique

Si un compte existe avec l'adresse e-mail fournie, un e-mail de réinitialisation du mot de passe vous sera envoyé. Si vous ne voyez pas d'e-mail, veuillez vérifier votre dossier de spam ou de courrier indésirable.

Pour plus d'assistance, contactez notre équipe d'assistance.

Veuillez vous connecter pour activer cette fonction

Pour accéder à cette fonctionnalité, vous devez être connecté à un compte OrthoEvidence actif. Veuillez vous connecter ou créer un compte d'essai GRATUIT.

Traduire le rapport ACE

OrthoEvidence utilise un service de traduction tiers pour rendre le contenu accessible dans plusieurs langues. Veuillez noter que même si tous les efforts sont faits pour assurer l'exactitude, les traductions ne sont pas toujours parfaites.

Comment citer ce document ACE Report

OrthoEvidence. Despite initial migration, Furlong Active stem comparable to Furlong HAC after 5 years. OE Journal. 2015;3(6):29. Available from: https://myorthoevidence.com/AceReport/Show/despite-initial-migration-furlong-active-stem-comparable-to-furlong-hac-after-5-years