OSTEOARTHRITIS
Despite initial migration, Furlong Active stem comparable to Furlong HAC after 5 years .

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2015;3(6):29 Acta Orthop. 2014 Dec;85(6):556-61.
المؤلفون المساهمون

E Weber M Sundberg G Flivik

50 patients with primary osteoarthritis undergoing total hip arthroplasty were randomized to receive either the Furlong Active femoral stem or the traditional Furlong HAC femoral stem. The purpose of this study was to evaluate the 3-D migratory patterns of the two stems using radiostereometry. Results after 3 months demonstrated significantly greater subsidence of the Furlong Active stem compared to the Furlong HAC stem, though subsidence of each stem plateaued after 3 months and displayed minor changes through to 5 years. Both groups displayed initial retroversion of the implant at 3 months, though minimal differences through to 5 years thereafter. Clinical outcome significantly improved in both groups over 5 years, with no significant differences detected between groups at 1, 2, or 5 year follow-up.


تفاصيل تمويل المنشور +
التمويل:
Industry funded
الراعي:
JRI Ltd.
التعارضات:
None disclosed

مخاطر التحيز

6٫5/10

معايير الإبلاغ

14/20

مؤشر الهشاشة

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

نعم = 1

غير مؤكد = 0.5

غير ذي صلة = 0

لا = 0

يقيّم تقييم معايير الإبلاغ الشفافية التي يبلغ بها المؤلفون عن الخصائص المنهجية والتجريبية للتجربة في المنشور. ينقسم التقييم إلى خمس فئات معروضة أدناه.

1/4

Randomization

2/4

Outcome Measurements

4/4

Inclusion / Exclusion

4/4

Therapy Description

3/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

مؤشر الهشاشة هو أداة تساعد في تفسير النتائج المهمة، وتوفر مقياسًا لقوة النتيجة. ويمثل مؤشر الهشاشة عدد الأحداث المتتالية التي يجب إضافتها إلى نتيجة ثنائية التفرع لجعل النتيجة غير مهمة. يمثل الرقم الصغير نتيجة أضعف ويمثل الرقم الكبير نتيجة أقوى.

لماذا كانت هناك حاجة لهذه الدراسة الآن؟

The Furlong HAC stem is a well-proven femoral implant that has recently undergone modifications to the lateral fin to prevent fracture of the greater trochanter during insertion. The new stem - the Furlong Active stem - is a shorter cone-shape, double-tapered, and has more rounded medial and lateral edges. Using radiostereometry, this study aimed to compare the migratory patterns of the Furlong HAC and Furlong Active stem as a surrogate for osseointegration and long-term survival of the arthroplasty.

ما هو سؤال البحث الرئيسي؟

Do design modifications to the Furlong HAC stem to give the Furlong Active stem affect the migratory behaviour, osseointegration and long-term result of total hip arthroplasty up to 5 years after surgery?

خصائص الدراسة +
Population:
50 patients (mean age: 63 years, mean BMI: 27, 27M/23F) with primary osteoarthritis scheduled for total hip arthroplasty. Patients who had unstable femoral anatomy as determined by radiographs and who were Charnley category C patients were excluded from the study. 45 patients (23M/20F) completed the study to completion at the 5 year endpoint. Surgery for both groups was performed using a posterolateral approach and 8-10 tantalum bone markers (diameter=0.8 mm). Apart from the femoral head used as a marker, all prostheses were pre-marked with tantalum markers at the tip, shoulder and medial collar of the stem.
Intervention:
Furlong Active group: Participants received the Furlong Active stem. Material was titanium alloy with a hydroxyapatite coating for uncemented implantation (n=25; 9M/16F) (Mean age: 62 years, range: 50-77 years)
Comparison:
Furlong HAC group: Participants received the Furlong HAC stem. Material was titanium alloy with a hydroxyapatite coating for uncemented implantation (n=25; 18M/7F) (Mean age: 63 years, range: 49-75 years)
Outcomes:
Radiostereometry (RSA) and conventional radiographic examinations were performed to analyze motion of the stem in relation to the tantalum markers implanted in the femur. Stem subsidence (Y-translation) and retroversion (Y-rotation) were primary effect variables. Self-administered Hip Disability and Osteoarthritis Outcome Score (HOOS) was also assessed.
Methods:
RCT: Single-centre
Time:
Radiostereometry (RSA) and conventional radiographic examinations were conducted on the first postoperative day, before weight-bearing and then at 3 months, 1,2 and 5 years (+/-5% time tolerance) postoperatively. Routine clinical follow-up occurred 3 months and 1, 2 and 5 months postoperatively. Hip Disability and Osteoarthritis Outcome Score was evaluated prior to surgery and at 1, 2 and 5 years.
ما هي النتائج المهمة؟
  • Significantly greater subsidence of the implant was noted in the Furlong Active group (0.99mm distally) compared to the Furlong HAC group (0.31mm distally) (p=0.0.46). Subsequent subsidence from 3 months to 5 years was not statistically significant in the Furlong Active (p=0.5) or HAC (p=0.06) stem.
  • There was one case of extreme subsidence reported in the Furlong Active group (6.5mm after 3 months) in a patient with BMI 41. Exclusion of this outlier case from the main analysis resulted in a non-significant difference in subsidence between groups (p>0.05).
  • Mean retroversion of the femoral stem was measured at 1.18deg in the Furlong Active group and 0.80deg in the Furlong HAC group after 3 months; the difference between groups was not statistically significant. (p=0.46). From 3 month to 5 years, no significant changes in mean retroversion were noted in either the Furlong Active (p=0.1) or HAC group (p=0.1).
  • Two patients, one in the Furlong HAC group (4.2 degrees) and one in the Furlong Active group (5.1 degree), demonstrated significant retroversion of the stem within the first 3 months after surgery. Between 3 months and 5 years, no further retroversion was noted in either patient.
  • No stem in either group was judged to be loose over the 5 year follow-up period.
  • HOOS scores improved significantly within both groups, without significant difference between groups at 1, 2, or 5 years after surgery.
  • Minor X- and Z-plane translation and rotation of the stem were noted in both groups, and differences were small between groups.
ما الذي يجب أن أتذكره أكثر؟

In total hip arthroplasty, subsidence of the uncemented Furlong Active femoral stem was significantly greater than the Furlong HAC stem in the first 3 months after surgery, although subsidence in both groups thereafter plateaued, with minimal subsidence occurring in either group from 3 months to 5 years. Both stems displayed initial retroversion of the femoral implant in the first 3 months, but minor changes thereafter to 5 years. No stem in either group was considered loose. Clinical outcome was similar between the Furlong Active and HAC stem throughout the 5-year follow-up.

كيف سيؤثر ذلك على رعاية مرضاي؟

Though subsidence of the Furlong Active stem was greater than that of the Furlong HAC stem, early migration did not appear to negatively impact 5-year outcome. Namely, osseointegration of the implant was observed to occur after 3 months, with minimal subsidence migration of either implant from 3 month to 5 years, and no loose components observed. Additionally, retroversion did not appear to substantial differ between stem designs. These findings suggest that similar results may be attained with the new Furlong Active stem relative the standard Furlong HAC stem, and that the new design modifications do not negatively affect integration of the component or clinical results. Nevertheless, additional research in a larger cohort of trials should be considered to confirm the the findings of the current study.

تنويه

هذا المحتوى الموجود في هذه الصفحة هو لأغراض إعلامية فقط وليس الغرض منه أن يكون بديلاً عن المشورة الطبية المتخصصة أو التشخيص أو العلاج. إذا كنت بحاجة إلى علاج طبي، اطلب دائمًا مشورة طبيبك أو اذهب إلى أقرب قسم طوارئ إليك. الآراء والمعتقدات ووجهات النظر التي يعبر عنها الأفراد في المحتوى الموجود في هذه الصفحة لا تعكس آراء ومعتقدات ووجهات نظر أورثوإيفيدنس.

0 من 4 مقالات مجانية شهرية غير مقفلة
لقد وصلت إلى الحد الأقصى لمشاهدة 4 من 4 مقالات مجانية هذا الشهر

ادخل إلى OrthoEvidence مقابل 1.99 دولارًا أمريكيًا في الأسبوع.

ابق على اتصال بأحدث الأدلة. قم بالإلغاء في أي وقت.
  • تقييمات نقدية لأحدث التجارب المعشاة ذات الشواهد عالية التأثير والمراجعات المنهجية في مجال جراحة العظام
  • الوصول إلى محتوى المدونة الصوتية OrthoEvidence، بما في ذلك التعاون مع مجلة جراحة العظام والمفاصل، والمقابلات مع الجراحين المعترف بهم دولياً، ومناقشات المائدة المستديرة حول أخبار وموضوعات جراحة العظام
  • الاشتراك في نشرة The Pulse، وهي نشرة إخبارية مبنية على الأدلة مرتين أسبوعيًا مصممة لمساعدتك على اتخاذ قرارات سريرية أفضل
Upgrade
Close Dialog
أهلاً بعودتك!
هل نسيت كلمة المرور؟
ابدأ تجربتك المجانية اليوم!

سيكون حسابك تابعًا لـ
ويتضمن وصولًا مجانيًا إلى OrthoEvidence


أو
هل نسيت كلمة المرور؟

أو
يرجى التحقق من بريدك الإلكتروني

في حالة وجود حساب بعنوان البريد الإلكتروني المقدم، سيتم إرسال بريد إلكتروني لإعادة تعيين كلمة المرور إليك. إذا كنت لا ترى بريدًا إلكترونيًا، يرجى التحقق من مجلد الرسائل غير المرغوب فيها أو البريد غير المرغوب فيه.

لمزيد من المساعدة اتصل بفريق الدعم لدينا.

يرجى تسجيل الدخول لتفعيل هذه الميزة

للوصول إلى هذه الميزة، يجب تسجيل الدخول إلى حساب OrthoEvidence نشط. يرجى تسجيل الدخول أو إنشاء حساب تجريبي مجاني.

ترجمة تقرير ACE

تستخدم OrthoEvidence خدمة ترجمة من طرف ثالث لإتاحة المحتوى بلغات متعددة. يُرجى ملاحظة أنه على الرغم من بذل كل جهد ممكن لضمان الدقة، إلا أن الترجمات قد لا تكون دائمًا مثالية.

كيفية الاستشهاد بهذا ACE Report

OrthoEvidence. Despite initial migration, Furlong Active stem comparable to Furlong HAC after 5 years. OE Journal. 2015;3(6):29. Available from: https://myorthoevidence.com/AceReport/Show/despite-initial-migration-furlong-active-stem-comparable-to-furlong-hac-after-5-years

نسخ الاقتباس
يرجى تسجيل الدخول لتفعيل هذه الميزة

للوصول إلى هذه الميزة، يجب تسجيل الدخول إلى حساب OrthoEvidence نشط. يرجى تسجيل الدخول أو إنشاء حساب تجريبي مجاني.

ميزة العضو المميز

للوصول إلى هذه الميزة، يجب عليك تسجيل الدخول إلى حساب OrthoEvidence المميز.

شارك هذا ACE Report