ACE Report Cover
THA: Stem migration and bone loss similar in cementless vs. cemented femoral components
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
Idioma
Download Download Download
Descargar
Cite this Report Cite this Report Cite this Report
Citar
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ Favoritos
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
Idioma
Download Download Download
Descargar
Cite this Report Cite this Report Cite this Report
Citar
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ Favoritos
ARTHROPLASTY
THA: Stem migration and bone loss similar in cementless vs. cemented femoral components .

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2013;1(6):26 Acta Orthop. 2012 Dec;83(6):592-8. doi: 10.3109/17453674.2012.742393. Epub 2012 Nov 1

51 patients, including 52 hips, undergoing treatment for osteoarthritis were randomized to receive either reverse hybrid (cemented polyethylene cup with cementless femoral stem) or conventional cemented total hip replacement (both components cemented) to determine if there was a difference in outcome between the techniques. Clinical outcome, wear, component migration and bone remodeling were recorded over 2 years postoperatively. The results showed that there were no major differences between the two procedures. This study indicated that the reverse hybrid procedure does not provide improved outcomes when compared to the conventional cemented method in total hip replacement.


Detalles de la financiación de la publicación +
Financiación:
Not Reported
Conflictos:
None disclosed

Riesgo de sesgo

6,5/10

Criterios de información

14/20

Índice de fragilidad

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

Sí = 1

Incierto = 0,5

No relevante = 0

No = 0

La evaluación de los criterios de información evalúa la transparencia con la que los autores informan de las características metodológicas y del ensayo dentro de la publicación. La evaluación se divide en cinco categorías que se presentan a continuación.

1/4

Randomization

3/4

Outcome Measurements

2/4

Inclusion / Exclusion

4/4

Therapy Description

4/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

El Índice de Fragilidad es una herramienta que ayuda en la interpretación de hallazgos significativos, proporcionando una medida de fuerza para un resultado. El Índice de Fragilidad representa el número de eventos consecutivos que es necesario añadir a un resultado dicotómico para que el hallazgo deje de ser significativo. Un número pequeño representa un hallazgo más débil y un número grande un hallazgo más fuerte.

¿Por qué se necesitaba ahora este estudio?

The Swedish Hip Arthroplasty and Norwegian Arthroplasty Registers have shown increased use of reverse hybrid technique (cemented polyethylene cup with cementless femoral stem) in total hip replacement. This is in part is due to the observed improvement in survivability of the cementless stems with respect to component loosening. However, there has been concern regarding polyethylene wear induced by hydroxyapatite (HA) in HA-coated implants, commonly seen in reverse hybrid total hip replacements. Also, there has not been a study conducted that has shown that the reverse hybrid technique is superior to conventional cemented total hip replacement, indicating the need for this study.

¿Cuál era la pregunta principal de la investigación?

Does the reverse hybrid total hip replacement technique provide superiority to conventional cemented total hip replacement with respect to clinical results, wear, remodeling of the bone, component migration, or component loosening after 2 years?

Características del estudio +
Población:
51 patients (52 hips) with osteoarthritis undergoing total hip replacement with cemented all-polyethylene cup. (n=43 total hips from 42 patients completed follow-up)
Intervención:
Reverse Hybrid Group: Patients underwent reverse hybrid total hip replacement with a cementless femoral stem, including proximal HA coating on top of the metal coating, tantalum balls attached at the tip, calcar region, and neck for radiostereometry analysis (RSA), and a 28-mm cobalt chrome head. (n=19 hips from 19 patients)
Comparación:
Cemented Group: Patients underwent conventional cemented total hip replacement with cemented femoral stem, with tantalum markers attached to the tip, calcar region and head (for RSA), and a 28-mm cobalt chrome head. A polyethylene distal cement restrictor was used and cement was pressurized prior to prostheses insertion. (n=24 hips from 23 patients)
Resultados:
Clinical outcomes were measured using Harris hip and Oxford hip scores for function and pain. Component wear and migration/rotation were evaluated using RSA. Bone remodeling and bone loss as a change in bone mineral density (BMD) were measured using dual-energy X-ray absorptiometry. Component loosening was analyzed via lucency in radiographs.
Métodos:
RCT, Single-centre
Tiempo:
RSA and BMD follow-up was conducted at 6 months, 1 and 2 years postoperatively. Follow-up of clinical outcomes (Harris and Oxford hip scores) and radiographic analysis were conducted at 2 years postoperatively.
¿Cuáles fueron los hallazgos importantes?
  • Similar improvements in Harris hip and Oxford hip scores were observed in both groups.
  • There were no significant differences observed between groups in the amount of wear in the x- (p=0.9), y- (p=0.9), and z-axes (p=0.08), and the vectorial resultant 3D wear (p=0.7).
  • Cup and stem component migration in any direction (x-, y-, or z-axes) were observed to not significantly differ between groups.
  • Rotation of stem component did not significantly differ between groups. Rotation of cup component in the y- and z-axes was not significantly different between groups. Rotation of the cup component in the x-axis was -0.24 degrees (95%CI -0.55 to 0.07) in the RH group and 0.13 degrees (95%CI -0.08 to 0.34) and in the cemented group, resulting in significant difference (p=0.03).
  • Bone remodeling around the cup was found to be similar between groups.
  • Significant bone loss in Gruen Zone 1 of the femur was observed in the RH group (18%, 95%CI 11-24%) when compared to the cemented group (1.4%, 95%CI -3.2-5.9%) (p<0.001). Bone loss in remaining Gruen zones (2-7) was not significant between groups.
  • There were no significant differences between groups for incidence of component loosening (p>0.05).
¿Qué es lo que más debo recordar?

The use of a cementless stem in total hip replacement provided similar results to a cemented stem with regards to Harris hip and Oxford hip scores, wear, component migration/rotation, bone remodelling, and bone loss after 2 years. Although significantly higher bone loss in Gruen Zone 1 was associated with the cementless stems, it is unlikely that it is an important clinical finding due to the long-term survivability of the cementless stems.

¿Cómo afectará esto al cuidado de mis pacientes?

There are no major differences between a cementless and a cemented femoral stem in total hip arthroplasty after 2 years. An extended study with a longer follow-up period is necessary to investigate any differences that may exist between the two techniques.

DESCARGO DE RESPONSABILIDAD

El contenido de esta página tiene únicamente fines informativos y no pretende sustituir el consejo, diagnóstico o tratamiento médico profesional. Si necesita tratamiento médico, busque siempre el consejo de su médico o acuda al servicio de urgencias más cercano. Las opiniones, creencias y puntos de vista expresados por las personas sobre el contenido que se encuentra en esta página no reflejan las opiniones, creencias y puntos de vista de OrthoEvidence.

0 de 4 artículos mensuales GRATIS desbloqueados
Ha alcanzado su límite de vistas de 4 artículos gratuitos este mes

Acceda a OrtoEvidencia por tan sólo 1,99 $ a la semana.

Manténgase conectado con las últimas pruebas. Cancele en cualquier momento.
  • Valoraciones críticas de los últimos ensayos controlados aleatorizados de gran impacto y revisiones sistemáticas en ortopedia
  • Acceso al contenido del podcast OrthoEvidence, que incluye colaboraciones con el Journal of Bone and Joint Surgery, entrevistas con cirujanos reconocidos internacionalmente y mesas redondas sobre noticias y temas ortopédicos
  • Suscripción a The Pulse, un boletín quincenal basado en la evidencia y diseñado para ayudarle a tomar mejores decisiones clínicas
Upgrade
Close Dialog
Bienvenido
¿Ha olvidado su contraseña?
Comience hoy mismo su prueba GRATUITA

Su cuenta estará afiliada a
e incluye acceso gratuito a OrthoEvidence


O
¿Olvidó su contraseña?

O
Compruebe su correo electrónico

Si existe una cuenta con la dirección de correo electrónico proporcionada, se le enviará un correo electrónico para restablecer la contraseña. Si no ve el correo electrónico, compruebe su carpeta de correo no deseado o spam.

Si necesita más ayuda póngase en contacto con nuestro equipo de asistencia.

Inicie sesión para activar esta función

Para acceder a esta función, debe iniciar sesión en una cuenta activa de OrthoEvidence. Por favor, inicie sesión o cree una cuenta de prueba GRATUITA.

Traducir Informe ACE

OrthoEvidence utiliza un servicio de traducción de terceros para que el contenido sea accesible en varios idiomas. Tenga en cuenta que, aunque se hace todo lo posible para garantizar la exactitud, las traducciones no siempre son perfectas.

Cómo citar esto ACE Report

OrthoEvidence. THA: Stem migration and bone loss similar in cementless vs. cemented femoral components. OE Journal. 2013;1(6):26. Available from: https://myorthoevidence.com/AceReport/Show/tha-stem-migration-and-bone-loss-similar-in-cementless-vs-cemented-femoral-components

Copiar cita
Inicie sesión para activar esta función

Para acceder a esta función, debe iniciar sesión en una cuenta activa de OrthoEvidence. Por favor, inicie sesión o cree una cuenta de prueba GRATUITA.

Función de miembro Premium

Para acceder a esta función, debe iniciar sesión en una cuenta Premium de OrthoEvidence.

Compartir ACE Report