ARTHROPLASTY
Similar migration & wear of an all-poly press-fit cup with and without screw stabilization .
Verified
This report has been verified by one or more authors of the original publication.

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2016;4(22):2 Acta Orthop. 2016 Aug;87(4):363-7
Autores contribuintes

MJ Minten PJ Heesterbeek M Spruit

36 patients who had undergone total hip arthroplasty and completed a 2-year follow-up of a previous randomized controlled trial were invited and included in this follow-up extension. Patients had received a press-fit, all-polyethylene acetabular cup, and had been randomized to placement with or without additional screw fixation. Radiostereometric analysis was performed to assess component migration and wear over 6.5-year follow-up, and the Harris Hip Score and Oxford Hip Score were used to assess clinical outcome. No significant differences were observed between groups in total 3-dimensional translation or rotation of the acetabular component. There were also no significant differences between groups in the total overall and corrected 3-dimensional wear or overall & corrected wear rates.


Detalhes do financiamento da publicação +
Financiamento:
Industry funded
Patrocinador:
Mathys Orthopaedics
Conflicts:
None disclosed

Risco de viés

6/10

Critérios de notificação

10/20

Índice de Fragilidade

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

Sim = 1

Incerto = 0,5

Não relevante = 0

Não = 0

A Avaliação dos Critérios de Relato avalia a transparência com que os autores relatam as caraterísticas metodológicas e do ensaio na publicação. A avaliação está dividida em cinco categorias que são apresentadas de seguida.

1/4

Randomization

3/4

Outcome Measurements

0/4

Inclusion / Exclusion

2/4

Therapy Description

4/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

O Índice de Fragilidade é uma ferramenta que auxilia na interpretação de achados significativos, fornecendo uma medida de força para um resultado. O Índice de Fragilidade representa o número de eventos consecutivos que precisam de ser adicionados a um resultado dicotómico para que o resultado deixe de ser significativo. Um número pequeno representa um resultado mais fraco e um número grande representa um resultado mais forte.

Porque é que este estudo era necessário agora?

Press-fit fixation of acetabular cups is often used in total hip arthroplasty. A debate has emerged surrounding the use of additional fixation screws in press-fit acetabular cups. While improved short-term stability may occur, there is concern that long-term stability could be compromised through various mechanisms, such as stress-shielding and wear. However, there is a current lack of evidence from randomized controlled trials to recommend either the use or avoidance of additional screw fixation with press-fit acetabular cups.

Qual era a principal questão de investigação?

In total hip arthroplasty, is there any significant difference in cup migration, cup wear, and clinical outcome of a cementless all-polyethylene acetabular cup with and without screw fixation, assessed over a 6.5-year follow-up?

Caraterísticas do estudo +
Population:
36 patients who had undergone total hip arthroplasty as part of a previous trial, and who had completed 2-year follow-up were included. All cases were performed through a posterolateral approach with standardized components across group: RM press-fit all-polyethylene acetabular cup (Mathys AG) with a cementless Spotorno femoral stem (Zimmer) and 32mm ceramic femoral head. Six tantalum markers were inserted into the acetabular component preoperatively, and a minimum of an additional six tantalum markers were inserted into the acetabular bone intraoperatively to facilitate radiosterometric analysis during follow-up.
Intervention:
Screw group: The acetabular component was additionally stabilized using two 4.0mm screws. (n=17; 15 analyzed)
Comparison:
No screw group: No additional stabilization of the acetabular component was used. (n=19; 12 analyzed)
Outcomes:
Radiostereometric analysis (RSA) was performed to calculate three-dimensional translation and rotation of the acetabular cup. RSA was additionally used to assess implant wear, defined as the translation of the femoral head in relation to the acetabular cup in all three planes. Total wear and wear rate were calculated, both as raw data as well as corrected for initial bedding-in over the first postoperative year. Additionally, the incidence of clinically relevant translation (>1mm) and clinically relevant rotation (>2 degrees) was assessed. Clinical outcome was assessed through the Harris Hip Score and the Oxford Hip Score. Device-related adverse events were also documented.
Methods:
RCT; single-center
Time:
Follow-up scheduled at 1, 2, and 6 years. Median follow-up was 6.4 and 6.3 years, respectively.
Quais foram os resultados importantes?
  • There was no significant difference between the screw group and no screw group in either total 3-dimensional translation (p=0.8) or total 3-dimensional rotation (p=0.2).
  • Between-group differences in individual planes were non-significant, with the exception of medial-lateral rotation: mean rotation was 0.33deg (+/-0.50) in the screw group and -0.61deg (+/-0.97) in the no screw group (p=0.004).
  • There were two cases of clinically relevant rotation of the acetabular component in each group. The same two cases in the no screw group also demonstrated clinically relevant translation of the acetabular component.
  • There was no significant difference in total wear over 6-year follow-up between the screw group (0.57+/-0.24mm) and the no screw group (0.53+/-0.20mm) (p=0.6). The only plane which demonstrated a significant difference between groups was medial-lateral wear (Screws: 0.19+/-0.22mm | No screws: -0.04+/-0.12mm | p=0.003).
  • Corrected wear analysis, which accounted for the large degree of wear within the first year after surgery, demonstrated similar results to the overall analysis.
  • Overall wear rate over 6 years did not significantly differ between the screw group (0.09+/-0.04 mm/year) and the no screw group (0.08+/-0.03 mm/year) (p=0.7). Corrected wear rate also did not significantly differ between groups (0.07+/-0.04 vs. 0.09+/-0.04 mm/year; p=0.3).
  • No significant difference between groups was noted either HHS (p=0.6) or OHS (p=0.6).
De que é que me devo lembrar mais?

In total hip arthroplasty with a press-fit acetabular cup, there were no significant differences in total 3-dimensional translation or rotation of the acetabular cup after 6.5 years, as assessed with RSA, with or without additional screw stabilization. There were also no significant differences in RSA-assessed component wear or clinical outcome after 6 years.

Como é que isto afectará o tratamento dos meus doentes?

The results of this study suggest that additional screw stabilization of cementless, press-fit all-polyethylene acetabular cup in total hip arthroplasty does not confer any significant mid-term advantage in migration, component wear, or clinical outcome over cups placed without additional stabilization.

AVISO LEGAL

O conteúdo desta página destina-se apenas a fins informativos e não pretende substituir o aconselhamento, diagnóstico ou tratamento médico profissional. Se necessitar de tratamento médico, procure sempre o conselho do seu médico ou dirija-se ao serviço de urgência mais próximo. As opiniões, crenças e pontos de vista expressos pelos indivíduos no conteúdo encontrado nesta página não reflectem as opiniões, crenças e pontos de vista da OrthoEvidence.

0 de 4 artigos mensais GRATUITOS desbloqueados
Atingiu o seu limite de 4 visualizações de artigos gratuitos este mês

Aceda à OrthoEvidence por apenas $1,99 por semana.

Mantenha-se ligado às últimas evidências. Cancele a qualquer altura.
  • Avaliações críticas dos mais recentes ensaios clínicos aleatórios de alto impacto e revisões sistemáticas em ortopedia
  • Acesso ao conteúdo do podcast OrthoEvidence, incluindo colaborações com o Journal of Bone and Joint Surgery, entrevistas com cirurgiões reconhecidos internacionalmente e discussões em mesa redonda sobre notícias e tópicos ortopédicos
  • Subscrição do The Pulse, um boletim informativo duas vezes por semana baseado em evidências, concebido para o ajudar a tomar melhores decisões clínicas
Upgrade
Bem-vindo de volta!
Esqueceu-se da palavra-passe?
Comece hoje o seu teste GRATUITO!

A sua conta será afiliada a
e inclui acesso gratuito ao OrthoEvidence


OU
Esqueceu-se da palavra-passe?

OU
Verifique o seu e-mail

Se existir uma conta com o endereço de e-mail fornecido, ser-lhe-á enviado um e-mail de reposição da palavra-passe. Se não vir uma mensagem de correio eletrónico, verifique a sua pasta de spam ou de lixo eletrónico.

Para obter mais assistência, contacte a nossa equipa de apoio.

Inicie sessão para ativar esta funcionalidade

Para aceder a esta funcionalidade, tem de iniciar sessão numa conta OrthoEvidence ativa. Inicie a sessão ou crie uma conta de avaliação GRATUITA.

Traduzir o Relatório ACE

A OrthoEvidence utiliza um serviço de tradução de terceiros para tornar o conteúdo acessível em vários idiomas. Tenha em atenção que, embora sejam feitos todos os esforços para garantir a exatidão, as traduções podem nem sempre ser perfeitas.

Como citar isto ACE Report

OrthoEvidence. Similar migration & wear of an all-poly press-fit cup with and without screw stabilization. OE Journal. 2016;4(22):2. Available from: https://myorthoevidence.com/AceReport/Show/similar-migration-wear-of-an-all-poly-press-fit-cup-with-and-without-screw-stabilization

Copie a citação
Inicie sessão para ativar esta funcionalidade

Para aceder a esta funcionalidade, tem de iniciar sessão numa conta OrthoEvidence ativa. Inicie a sessão ou crie uma conta de avaliação GRATUITA.

Funcionalidade de Membro Premium

Para aceder a esta funcionalidade, tem de ter sessão iniciada numa conta Premium da OrthoEvidence.

Partilhe isto ACE Report