OTA 2013: Locking plates comparable to nonlocking plates for high-energy pilon fracture .
OrthoEvidence Journal (OE Journal) - ACE Report
OE Journal. 2025;13(3):858 patients with high-energy pilon fractures were randomized to evaluate the efficacy of locking plates against convention nonlocking plates, over 1 year. Radiographic data was evaluated for loss of reduction using Mortise/ AP and lateral radiographs of the ankle, and functional outcomes were assessed using ankle hindfoot scores, and Short Musculoskeletal Function Assessment (SMFA). The evidence presented in this study demonstrates that both treatments perform comparably in each outcome measurement, but the authors of the study indicate that the results may suffer from type II error due to the low incidence of reduction loss.
Informe de la conferencia
Este informe de la ACE es un resumen de una presentación o resumen de conferencia. La información proporcionada ha limitado la capacidad de ofrecer una evaluación precisa del riesgo de sesgo o de la calidad general. Por favor, interprete los resultados con precaución ya que los ensayos pueden estar en curso y se pueden haber presentado resultados seleccionados.
¿Por qué se necesitaba ahora este estudio?
High-energy pilon fractures are conventionally treated with nonlocking plates. This study was needed to evaluate and compare the clinical and radiographic outcomes associated with the use of locking plates against nonlocking plates for the management of high-energy pilon fractures.
¿Cuál era la pregunta principal de la investigación?
Do locking plates present comparable clinical and radiographic outcomes to nonlocking plates for the treatment of high-energy pilon fractures, assessed over 1 year?
- 2 of 15 (13%) of the fractures in the nonlocking group demonstrated loss of reduction >5 degrees, compared to 3 of 19 (16%) in the lock group (p= 0.999)
- No soft-tissue complications requiring surgical intervention were observed
- 1 patient in the lock group presented with deep infection, 2 with nonunions, and 2 with hardware failures. 1 patient in the nonlock group presented with hardware failure and 1 infected nonunion
- Average follow-up in 8 patients in the lock group, and 10 in the nonlock group, was reported at 35.6 SD 16.0 months.
- No significant differences between the lock and nonlock groups for ankle hindfoot scores: 71.75 SD 71.75 versus 66.1 SD 23.8 respectively (p= 0.625)
- No significant difference between groups was apparent for SMFA-BI (bother index) scores with 7175 SD 25.4 in the lock group compared to 66.1 SD 23.8 in the nonlock group (p= 0.625)
- SMFA-FI (function index) analysis between the lock and nonlock groups revealed no significant difference with a score of 32.9 SD 36.2 in the lock group compared to 25.7 SD 20.2 in the nonlock group (p= 0.587)
¿Qué es lo que más debo recordar?
Both the locking plate and the nonlocking plates performed comparably for all evaluated outcomes: loss of reduction and radiologic analyses.
¿Cómo afectará esto al cuidado de mis pacientes?
The evidence presented in this study demonstrates that both locking and nonlocking plates are comparable interventions for high-energy pilon fractures. The results presented in this study may be subject to type II error as indicated by the low incidence of reduction loss.
DESCARGO DE RESPONSABILIDAD
El contenido de esta página tiene únicamente fines informativos y no pretende sustituir el consejo, diagnóstico o tratamiento médico profesional. Si necesita tratamiento médico, busque siempre el consejo de su médico o acuda al servicio de urgencias más cercano. Las opiniones, creencias y puntos de vista expresados por las personas sobre el contenido que se encuentra en esta página no reflejan las opiniones, creencias y puntos de vista de OrthoEvidence.