Implant migration reduced in hydroxyapatite coated versus uncoated components .
This report has been verified
by one or more authors of the
original publication.
OrthoEvidence Journal (OE Journal) - ACE Report
OE Journal. 2013;1(1):6 Acta Orthop. 2012 Apr;83(2):135-41. doi: 10.3109/17453674.2012.665330. Epub 2012 Feb 13Exclusive Author Interview
Dr. B.G. Pijls speaks on the long term beneficial effect of hydroxyapatite in TKA.
48 patients (68 knees) scheduled for total knee arthroplasty were randomized to receive a cemented, cementless non-coated, or cementless hydroxyapatite (HA)-coated tibial component to determine the effect of HA on long-term migration of the prosthesis. Radiostereometric analyses (RSA), Knee Society Scores, Hospital for Special Surgery scores, and radiographic evaluation were recorded over 16 years. The results indicated that migration of HA-coated components was improved compared to non-coated components, but increased in comparison to cemented components.
Was the allocation sequence adequately generated?
Was allocation adequately concealed?
Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?
Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?
Blinding Patients: Was knowledge of the allocated interventions adequately prevented?
Was loss to follow-up (missing outcome data) infrequent?
Are reports of the study free of suggestion of selective outcome reporting?
Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?
Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?
Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?
Sim = 1
Incerto = 0,5
Não relevante = 0
Não = 0
A Avaliação dos Critérios de Relato avalia a transparência com que os autores relatam as caraterísticas metodológicas e do ensaio na publicação. A avaliação está dividida em cinco categorias que são apresentadas de seguida.
4/4
Randomization
4/4
Outcome Measurements
2/4
Inclusion / Exclusion
2/4
Therapy Description
3/4
Statistics
Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65
O Índice de Fragilidade é uma ferramenta que auxilia na interpretação de achados significativos, fornecendo uma medida de força para um resultado. O Índice de Fragilidade representa o número de eventos consecutivos que precisam de ser adicionados a um resultado dicotómico para que o resultado deixe de ser significativo. Um número pequeno representa um resultado mais fraco e um número grande representa um resultado mais forte.
Porque é que este estudo era necessário agora?
Although the beneficial early fixation properties of hydroxyapatite (HA) coatings on cementless prosthetic components in total knee arthroplasty have been extensively studied and researched, reports of the long-term migration of HA-coated components are limited. Moreover, complications of delamination and third-body wear associated with the HA-coating have been reported in total hip arthroplasty, which could cause long-term problems.
Qual era a principal questão de investigação?
Does HA-coating maintain fixation properties and reduce migration of components in total knee arthroplasty over the long-term (11 to 16 years) when compared to uncoated components and cemented components?
- Migration of the HA-coated tibial components was less than the non-coated components (MD 0.39mm, 95%CI 0.16-0.62)
- HA-coated components were observed to have significantly greater migration compared to cemented components (MD 0.61mm, 95%CI 0.42-0.80, p<0.001)
- HA-coating decreased subsidence by 0.26mm (95%CI 0.10-0.42) and external rotation by 0.47 degrees (95%CI 0.27-0.67) when compared to non-coated components.
- There was no significant difference observed between groups in clinical outcomes of KSS and HSS at 10 years postoperatively.
- The groups did not significantly differ in radiographic assessments of FTA, alpha, beta, and delta angles of components.
- 49 of the operated knees underwent TKA as a result of rheumatoid arthritis, 18 due to osteoarthritis, and 1 because of sequelae after septic arthritis.
- Longitudinal follow-up of TKA with RSA allowed for early detection of secondary loosening.
De que é que me devo lembrar mais?
The use of HA-coating on cementless components for total knee arthroplasty decreased prosthesis migration significantly, when compared to non-coated cementless components. However, an increase in migration was observed when HA-coated components were compared to cemented components.
Como é que isto afectará o tratamento dos meus doentes?
HA-coating of cementless tibial components can reduce long-term migration compared to non-coated cementless components. However, cemented components appear to provide the greatest reduction in migration. Further research is required to confirm these findings in studies using larger sample sizes and a more balanced population (equal number of patients requiring treatment due to rheumatoid arthritis, osteoarthritis and other disorders).
AVISO LEGAL
O conteúdo desta página destina-se apenas a fins informativos e não pretende substituir o aconselhamento, diagnóstico ou tratamento médico profissional. Se necessitar de tratamento médico, procure sempre o conselho do seu médico ou dirija-se ao serviço de urgência mais próximo. As opiniões, crenças e pontos de vista expressos pelos indivíduos no conteúdo encontrado nesta página não reflectem as opiniões, crenças e pontos de vista da OrthoEvidence.