ACE Report Cover
Implant migration reduced in hydroxyapatite coated versus uncoated components
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
اللغة
Download Download Download
تحميل
Cite this Report Cite this Report Cite this Report
اقتباس
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ المفضلة
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
اللغة
Download Download Download
تحميل
Cite this Report Cite this Report Cite this Report
اقتباس
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ المفضلة
ARTHROPLASTY
Implant migration reduced in hydroxyapatite coated versus uncoated components .
Verified
This report has been verified by one or more authors of the original publication.

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2013;1(1):6 Acta Orthop. 2012 Apr;83(2):135-41. doi: 10.3109/17453674.2012.665330. Epub 2012 Feb 13
المؤلفون المساهمون

BG Pijls ER Valstar BL Kaptein M Fiocco RG Nelissen

Exclusive Author Interview

Dr. B.G. Pijls speaks on the long term beneficial effect of hydroxyapatite in TKA.

48 patients (68 knees) scheduled for total knee arthroplasty were randomized to receive a cemented, cementless non-coated, or cementless hydroxyapatite (HA)-coated tibial component to determine the effect of HA on long-term migration of the prosthesis. Radiostereometric analyses (RSA), Knee Society Scores, Hospital for Special Surgery scores, and radiographic evaluation were recorded over 16 years. The results indicated that migration of HA-coated components was improved compared to non-coated components, but increased in comparison to cemented components.


تفاصيل تمويل المنشور +
التمويل:
Non-Industry funded
الراعي:
Atlantic Canada Opportunities Agency, Dutch Arthritis Association
التعارضات:
None disclosed

مخاطر التحيز

7/10

معايير الإبلاغ

15/20

مؤشر الهشاشة

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

نعم = 1

غير مؤكد = 0.5

غير ذي صلة = 0

لا = 0

يقيّم تقييم معايير الإبلاغ الشفافية التي يبلغ بها المؤلفون عن الخصائص المنهجية والتجريبية للتجربة في المنشور. ينقسم التقييم إلى خمس فئات معروضة أدناه.

4/4

Randomization

4/4

Outcome Measurements

2/4

Inclusion / Exclusion

2/4

Therapy Description

3/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

مؤشر الهشاشة هو أداة تساعد في تفسير النتائج المهمة، وتوفر مقياسًا لقوة النتيجة. ويمثل مؤشر الهشاشة عدد الأحداث المتتالية التي يجب إضافتها إلى نتيجة ثنائية التفرع لجعل النتيجة غير مهمة. يمثل الرقم الصغير نتيجة أضعف ويمثل الرقم الكبير نتيجة أقوى.

لماذا كانت هناك حاجة لهذه الدراسة الآن؟

Although the beneficial early fixation properties of hydroxyapatite (HA) coatings on cementless prosthetic components in total knee arthroplasty have been extensively studied and researched, reports of the long-term migration of HA-coated components are limited. Moreover, complications of delamination and third-body wear associated with the HA-coating have been reported in total hip arthroplasty, which could cause long-term problems.

ما هو سؤال البحث الرئيسي؟

Does HA-coating maintain fixation properties and reduce migration of components in total knee arthroplasty over the long-term (11 to 16 years) when compared to uncoated components and cemented components?

خصائص الدراسة +
Population:
48 patients undergoing total knee arthroplasty for treatment of osteoarthritis or rheumatoid arthritis. TKAs were performed with standard midline incision and medial parapatellar arthrotomy. Tantalum markers were inserted into components, and all inserts were made of ultra-high-molecular-weight-polyethylene, gamma radiation sterilized and machined from GUR 415 resin with calcium stearate. (n=68 total knees)
Intervention:
HA-coated Group: Patients received cementless, HA-coated tibial components. (n=24)
Comparison:
Uncoated Group: Patients received cementless uncoated tibial components. (n=20) Cemented Group: Patients received cemented tibial components. (n=24)
Outcomes:
Primary outcome was migration of component, assessed by RSA. Secondary outcome measurements were clinical assessments (KSS and HSS) and radiographic evaluation for component loosening (conventional weight-bearing radiographs of femoral-tibial alignment (FTA angle), frontal angle of femoral component (alpha), and frontal (beta) and sagittal (delta) angles of the tibial component)
Methods:
RCT, Double-/Single-Blind (assessors blinded for HA comparison in cementless components, but not for comparison between cemented and cementless components), Prospective
Time:
Follow-up of RSA and clinical outcomes was conducted at 3 weeks, 6 weeks, 3 months, 6 months, 1 year, and annually thereafter. Radiographic evaluation occurred 2, 5, 10 and 15 years postoperatively.
ما هي النتائج المهمة؟
  • Migration of the HA-coated tibial components was less than the non-coated components (MD 0.39mm, 95%CI 0.16-0.62)
  • HA-coated components were observed to have significantly greater migration compared to cemented components (MD 0.61mm, 95%CI 0.42-0.80, p<0.001)
  • HA-coating decreased subsidence by 0.26mm (95%CI 0.10-0.42) and external rotation by 0.47 degrees (95%CI 0.27-0.67) when compared to non-coated components.
  • There was no significant difference observed between groups in clinical outcomes of KSS and HSS at 10 years postoperatively.
  • The groups did not significantly differ in radiographic assessments of FTA, alpha, beta, and delta angles of components.
  • 49 of the operated knees underwent TKA as a result of rheumatoid arthritis, 18 due to osteoarthritis, and 1 because of sequelae after septic arthritis.
  • Longitudinal follow-up of TKA with RSA allowed for early detection of secondary loosening.
ما الذي يجب أن أتذكره أكثر؟

The use of HA-coating on cementless components for total knee arthroplasty decreased prosthesis migration significantly, when compared to non-coated cementless components. However, an increase in migration was observed when HA-coated components were compared to cemented components.

كيف سيؤثر ذلك على رعاية مرضاي؟

HA-coating of cementless tibial components can reduce long-term migration compared to non-coated cementless components. However, cemented components appear to provide the greatest reduction in migration. Further research is required to confirm these findings in studies using larger sample sizes and a more balanced population (equal number of patients requiring treatment due to rheumatoid arthritis, osteoarthritis and other disorders).

تنويه

هذا المحتوى الموجود في هذه الصفحة هو لأغراض إعلامية فقط وليس الغرض منه أن يكون بديلاً عن المشورة الطبية المتخصصة أو التشخيص أو العلاج. إذا كنت بحاجة إلى علاج طبي، اطلب دائمًا مشورة طبيبك أو اذهب إلى أقرب قسم طوارئ إليك. الآراء والمعتقدات ووجهات النظر التي يعبر عنها الأفراد في المحتوى الموجود في هذه الصفحة لا تعكس آراء ومعتقدات ووجهات نظر أورثوإيفيدنس.

0 من 4 مقالات مجانية شهرية غير مقفلة
لقد وصلت إلى الحد الأقصى لمشاهدة 4 من 4 مقالات مجانية هذا الشهر

ادخل إلى OrthoEvidence مقابل 1.99 دولارًا أمريكيًا في الأسبوع.

ابق على اتصال بأحدث الأدلة. قم بالإلغاء في أي وقت.
  • تقييمات نقدية لأحدث التجارب المعشاة ذات الشواهد عالية التأثير والمراجعات المنهجية في مجال جراحة العظام
  • الوصول إلى محتوى المدونة الصوتية OrthoEvidence، بما في ذلك التعاون مع مجلة جراحة العظام والمفاصل، والمقابلات مع الجراحين المعترف بهم دولياً، ومناقشات المائدة المستديرة حول أخبار وموضوعات جراحة العظام
  • الاشتراك في نشرة The Pulse، وهي نشرة إخبارية مبنية على الأدلة مرتين أسبوعيًا مصممة لمساعدتك على اتخاذ قرارات سريرية أفضل
Upgrade
Close Dialog
أهلاً بعودتك!
هل نسيت كلمة المرور؟
ابدأ تجربتك المجانية اليوم!

سيكون حسابك تابعًا لـ
ويتضمن وصولًا مجانيًا إلى OrthoEvidence


أو
هل نسيت كلمة المرور؟

أو
يرجى التحقق من بريدك الإلكتروني

في حالة وجود حساب بعنوان البريد الإلكتروني المقدم، سيتم إرسال بريد إلكتروني لإعادة تعيين كلمة المرور إليك. إذا كنت لا ترى بريدًا إلكترونيًا، يرجى التحقق من مجلد الرسائل غير المرغوب فيها أو البريد غير المرغوب فيه.

لمزيد من المساعدة اتصل بفريق الدعم لدينا.

يرجى تسجيل الدخول لتفعيل هذه الميزة

للوصول إلى هذه الميزة، يجب تسجيل الدخول إلى حساب OrthoEvidence نشط. يرجى تسجيل الدخول أو إنشاء حساب تجريبي مجاني.

ترجمة تقرير ACE

تستخدم OrthoEvidence خدمة ترجمة من طرف ثالث لإتاحة المحتوى بلغات متعددة. يُرجى ملاحظة أنه على الرغم من بذل كل جهد ممكن لضمان الدقة، إلا أن الترجمات قد لا تكون دائمًا مثالية.

كيفية الاستشهاد بهذا ACE Report

OrthoEvidence. Implant migration reduced in hydroxyapatite coated versus uncoated components. OE Journal. 2013;1(1):6. Available from: https://myorthoevidence.com/AceReport/Show/implant-migration-reduced-in-hydroxyapatite-coated-versus-uncoated-components

نسخ الاقتباس
يرجى تسجيل الدخول لتفعيل هذه الميزة

للوصول إلى هذه الميزة، يجب تسجيل الدخول إلى حساب OrthoEvidence نشط. يرجى تسجيل الدخول أو إنشاء حساب تجريبي مجاني.

ميزة العضو المميز

للوصول إلى هذه الميزة، يجب عليك تسجيل الدخول إلى حساب OrthoEvidence المميز.

شارك هذا ACE Report