No difference between Cemex Rx and Palacos R for THR .
OrthoEvidence Journal (OE Journal) - ACE Report
OE Journal. 2013;1(6):15 Acta Orthop. 2012 Dec;83(6):604-8. doi: 10.3109/17453674.2012.742392. Epub 2012 Nov 1.44 patients undergoing total hip replacement were randomized to receive a Lubinus SPII titanium-aluminum-vanadium stem cemented either with the new Cemex Rx bone cement or with Palacos R bone cement. The 10 year results of the study indicated that Cemex bone cement performs just as well as Palacos R bone cement.
Was the allocation sequence adequately generated?
Was allocation adequately concealed?
Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?
Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?
Blinding Patients: Was knowledge of the allocated interventions adequately prevented?
Was loss to follow-up (missing outcome data) infrequent?
Are reports of the study free of suggestion of selective outcome reporting?
Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?
Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?
Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?
Oui = 1
Incertain = 0,5
Non pertinent = 0
Non = 0
L'évaluation des critères de rapport permet d'évaluer la transparence avec laquelle les auteurs rapportent les caractéristiques méthodologiques et les caractéristiques de l'essai dans la publication. L'évaluation est divisée en cinq catégories qui sont présentées ci-dessous.
2/4
Randomization
3/4
Outcome Measurements
1/4
Inclusion / Exclusion
4/4
Therapy Description
4/4
Statistics
Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65
L'indice de fragilité est un outil qui aide à l'interprétation des résultats significatifs, en fournissant une mesure de la force d'un résultat. L'indice de fragilité représente le nombre d'événements consécutifs qui doivent être ajoutés à un résultat dichotomique pour que le résultat ne soit plus significatif. Un petit nombre représente un résultat plus faible et un grand nombre un résultat plus fort.
Pourquoi cette étude était-elle nécessaire maintenant ?
Enhancements in cementing techniques during cemented total hip arthroplasty have reduced the incidence of early loosening and delayed revision time. However, the curing temperature during the cementing process is believed to cause bone and cell necrosis, which leads to membrane formation and aseptic loosening, a cause for revision. Cemex was designed to reduce the curing temperature by removing the smallest sizes of particles in the polymer powder, which would lead to a smaller surface area reacting with monomers. But, laboratory tests have shown that the curing temperature of Cemex is no different that the curing temperature of conventional Palacos R. This study aimed to determine aimed to examine the clinical properties of Cemex and compare the results with those of Palacos R.
Quelle était la principale question de recherche ?
Does Cemex Rx bone cement reduce femoral stem migration and rotation while improving hip pain and function when compared to Palacos R bone cement among total hip replacement patients, 10 years after surgery?
- The magnitudes of rotation into flexion/extension and retroversion were significantly larger among the patients that received Palacos bone cement (p=0.005 and 0.01, respectively) but there was no significant difference in rotation around the sagittal axis as most implants were tilted into varus (p=0.7).
- The magnitude of subsidence was similar among the patients that received Palacos bone cement and the patients that received Cemex bone cement (p=0.7).
- There was no significant differences in migration of the femoral component in relation to the cement mantle between the treatment groups. Most migration occurred between the stem and cement mantle in both groups.
- Radiolucent lines along the stem were found in Gruen zones 2 and 6 in 22 cases. The proportion of radiolucent lines in the total interface was 40% for the patients that received Palacos and 50% for the patients that received Cemex.
- There was no significant difference in Harris hip scores and pain scores between the patients that received Cemex bone cement and Palacos bone cement.
De quoi dois-je me souvenir en priorité ?
There was no significant difference in rotation around the sagital axis, the magnitude of subsidence, migration of the femoral component, radiolucent lines, hip function, and pain between the two treatment groups.
Comment cela affectera-t-il les soins prodigués à mes patients ?
The study suggests that Cemex Rx bone cement will preform similarly to Palacos R bone cement when used for total hip replacements. Further research trials are needed to confirm these findings.
AVIS DE NON-RESPONSABILITÉ
Le contenu de cette page est fourni à titre d'information uniquement et n'est pas destiné à remplacer un avis médical, un diagnostic ou un traitement professionnel. Si vous avez besoin d'un traitement médical, demandez toujours l'avis de votre médecin ou rendez-vous au service des urgences le plus proche. Les opinions, croyances et points de vue exprimés par les individus sur le contenu de cette page ne reflètent pas les opinions, croyances et points de vue d'OrthoEvidence.