Systemic cobalt release after metal-on-metal total hip arthroplasty: 5 year follow-up .
Serum cobalt levels after metal-on-metal total hip arthroplasty
J Bone Joint Surg Am. 2003 Nov;85-A(11):2168-73100 patients undergoing a total hip arthroplasty without cement for unilateral osteoarthritis or unilateral osteonecrosis were randomized to be treated with a metal-on-metal articulation or with a ceramic-on-polyethylene bearing. Blood samples were taken multiple times over a 5 year period and serum cobalt concentrations measured. In the metal articulation group, median cobalt levels were 1 μg/L at 1 year postoperatively and 0.7 μg/L at 5 years. The median cobalt levels in the ceramic-on-polyethylene group were below detection limits at all times. The serum cobalt concentrations did not reflect a so-called run-in wear period of the metal-on-metal articulations.
Vollständigen ACE-Bericht freischalten
Sie haben Zugang zu 4 weiteren KOSTENLOSEN Artikeln in diesem Monat.
Klicken Sie unten, um diese ACE Reports freizuschalten und anzusehen
Jetzt freischalten
Kritische Beurteilungen der neuesten, hochwirksamen randomisierten kontrollierten Studien und systematischen Übersichten in der Orthopädie
Zugang zu OrthoEvidence-Podcast-Inhalten, einschließlich Kooperationen mit dem Journal of Bone and Joint Surgery, Interviews mit international anerkannten Chirurgen und Diskussionsrunden zu orthopädischen Neuigkeiten und Themen
Abonnement von The Pulse, einem zweimal wöchentlich erscheinenden evidenzbasierten Newsletter, der Ihnen helfen soll, bessere klinische Entscheidungen zu treffen
Exklusiver Zugang zu Originalartikeln, einschließlich eigener systematischer Übersichten, sowie zu Artikeln über Methoden der Gesundheitsforschung und aktuelle orthopädische Themen