Migration similar between cemented and uncemented cups in trapeziometacarpal arthroplasty .
This report has been verified
by one or more authors of the
original publication.
OrthoEvidence Journal (OE Journal) - ACE Report
OE Journal. 2013;1(3):33 Acta Orthop. 2013 Feb;84(1):98-105. doi: 10.3109/17453674.2013.765625. Epub 2013 Jan 2328 patients (32 hands) were randomized to evaluate the effectiveness of cemented versus uncemented trapezium cups in total arthroplasty of the trapeziometacarpal joint. Patients were allocated to receive either uncemented Elektra screwed trapezium cups or cemented DLC all-polythelene trapezium cups, and underwent radiostereometric analysis and clinical evaluation for 2 years. Results at 2 years indicated that there were no significant differences between groups for trapezium component migration, and clinical outcomes were similar.
Was the allocation sequence adequately generated?
Was allocation adequately concealed?
Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?
Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?
Blinding Patients: Was knowledge of the allocated interventions adequately prevented?
Was loss to follow-up (missing outcome data) infrequent?
Are reports of the study free of suggestion of selective outcome reporting?
Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?
Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?
Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?
Ja = 1
Ungewiss = 0.5
Nicht relevant = 0
Nein = 0
Die Bewertung der Berichtskriterien bewertet die Transparenz, mit der die Autoren die methodischen und studienspezifischen Merkmale der Studie in der Veröffentlichung angeben. Die Bewertung ist in fünf Kategorien unterteilt, die im Folgenden vorgestellt werden.
3/4
Aleatorización
3/4
Medición de resultados
4/4
Inclusión / exclusión
4/4
Descripción de la terapia
4/4
Estadísticas
Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65
Der Fragilitätsindex ist ein Instrument, das bei der Interpretation signifikanter Ergebnisse hilft und ein Maß für die Stärke eines Ergebnisses liefert. Der Fragilitätsindex gibt die Anzahl der aufeinanderfolgenden Ereignisse an, die zu einem dichotomen Ergebnis hinzugefügt werden müssen, damit das Ergebnis nicht mehr signifikant ist. Eine kleine Zahl steht für ein schwächeres Ergebnis und eine große Zahl für ein stärkeres Ergebnis.
Warum wurde diese Studie jetzt benötigt?
High rates of component loosening - both cemented and uncemented - have been reported in total arthroplasty for trapeziometacarpal joint osteoarthritis, primarily with the trapezium cup. The gold standard implants are considered to be cemented all-polyethylene trapezium cups, yet some studies have found comparable outcomes with the uncemented Elektra hydroxyapatite(HA)-coated trapezium screw cup, with clinical importance of the results undetermined. Therefore, this study was designed to compare the clinical fixation of the two implants.
Was war die wichtigste Forschungsfrage?
Did uncemented Elektra screw cups (inserted without threading of the bone) provide similar clinical fixation to cemented DLC trapezium cups when measured after 2 years?
- 2 year migration (total translation) of uncemented Elektra cups was 0.80mm (95%CI -0.56 to 2.2) and 0.36mm (95%CI 0.07-0.64) in DLC all-poly cups, resulting in a statistically insignificant difference of -0.44mm (95%CI -1.7 to 0.86, p=0.3).
- 2 uncemented Elektra cups and 1 cemented DLC cup were classified as loose (total translation > 1.0mm).
- Analysis was also conducted excluding 2 outliers (1 from each group), and found 2-year total translation of 0.19mm (95%CI 0.07-0.31) in uncemented cups and 0.24mm (95%CI 0.16-0.31) in cemented cups.
- No significant differences were observed between groups in grip strength, VAS scores at rest or during activity, or DASH scores at any time throughout the duration of study (P > 0.05).
Was sollte ich mir besonders merken?
Component migration and clinical outcome were similar between uncemented Elektra HA-coated screwed trapezium cup or cemented DLC all-polyethylene trapezium cup 2 years following total arthroplasty of the trapeziometacarpal joint.
Wie wird sich dies auf die Behandlung meiner Patienten auswirken?
The results of this study suggest that either cemented or uncemented cups in trapeziometacarpal arthroplasty provide comparably good fixation properties. However, investigation into the long-term survival and outcomes of uncemented cups should be considered, as well as the implications of metal-on-metal articulation in total trapeziometacarpal arthroplasty.
HAFTUNGSAUSSCHLUSS
Der Inhalt dieser Seite dient nur zu Informationszwecken und ist nicht als Ersatz für professionelle medizinische Beratung, Diagnose oder Behandlung gedacht. Wenn Sie eine medizinische Behandlung benötigen, wenden Sie sich immer an Ihren Arzt oder suchen Sie die nächstgelegene Notaufnahme auf. Die Meinungen, Überzeugungen und Standpunkte, die von den Personen auf dieser Seite geäußert werden, spiegeln nicht die Meinungen, Überzeugungen und Standpunkte von OrthoEvidence wider.