ACE Report Cover
Metasul metal-on-metal THRs have high rate of revision and incidences of radiolucent lines
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
Lingua
Download Download Download
Scaricare
Cite this Report Cite this Report Cite this Report
Citare
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ Preferiti
Translate this  ACE Report Translate this  ACE Report Translate this  ACE Report
Lingua
Download Download Download
Scaricare
Cite this Report Cite this Report Cite this Report
Citare
Add to Favorites Add to Favorites Add to Favorites Remove from Favorites Remove from Favorites Remove from Favorites
+ Preferiti
ARTHROPLASTY
Metasul metal-on-metal THRs have high rate of revision and incidences of radiolucent lines .
Verified
This report has been verified by one or more authors of the original publication.

OrthoEvidence Journal (OE Journal) - ACE Report

OE Journal. 2013;1(16):1 Bone Joint J. 2013 Jul;95-B(7):881-6. doi: 10.1302/0301-620X.95B7.31067

399 hips undergoing total hip replacement were randomized to receive either Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP), or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered polished femoral component (MS-30). The study aimed to compare the three types of bearings in clinical and radiological outcomes. The results taken at 2,5,and 7 years revealed that MoM group had the lowest Harris Hips scores, higher rates of revision, and higher incidence of radiolucent lines in unrevised hips.


Dettagli sul finanziamento della pubblicazione +
Finanziamento:
Non-funded
Conflitti:
None disclosed

Rischio di pregiudizio

5/10

Criteri di segnalazione

16/20

Indice di fragilità

N/A

Was the allocation sequence adequately generated?

Was allocation adequately concealed?

Blinding Treatment Providers: Was knowledge of the allocated interventions adequately prevented?

Blinding Outcome Assessors: Was knowledge of the allocated interventions adequately prevented?

Blinding Patients: Was knowledge of the allocated interventions adequately prevented?

Was loss to follow-up (missing outcome data) infrequent?

Are reports of the study free of suggestion of selective outcome reporting?

Were outcomes objective, patient-important and assessed in a manner to limit bias (ie. duplicate assessors, Independent assessors)?

Was the sample size sufficiently large to assure a balance of prognosis and sufficiently large number of outcome events?

Was investigator expertise/experience with both treatment and control techniques likely the same (ie.were criteria for surgeon participation/expertise provided)?

Sì = 1

Incerto = 0,5

Non rilevante = 0

No = 0

La valutazione dei criteri di segnalazione valuta la trasparenza con cui gli autori riportano le caratteristiche metodologiche e sperimentali dello studio all'interno della pubblicazione. La valutazione è suddivisa in cinque categorie che vengono presentate di seguito.

3/4

Randomization

2/4

Outcome Measurements

4/4

Inclusion / Exclusion

4/4

Therapy Description

3/4

Statistics

Detsky AS, Naylor CD, O'Rourke K, McGeer AJ, L'Abbé KA. J Clin Epidemiol. 1992;45:255-65

L'Indice di Fragilità è uno strumento che aiuta l'interpretazione dei risultati significativi, fornendo una misura della forza di un risultato. L'Indice di Fragilità rappresenta il numero di eventi consecutivi che devono essere aggiunti a un risultato dicotomico per rendere il risultato non più significativo. Un numero piccolo rappresenta un risultato più debole, mentre un numero grande rappresenta un risultato più forte.

Perché questo studio era necessario ora?

Metasul metal-on-metal (MoM) articulation has shown lower wear rates than metal-on-polyethylene (MoP) bearings in Total Hip Replacement (THR), especially when used in cementless fixation. However, some evidence has reported that cemented MoM articualtions using 28 mm bearing have high rates of radiolucent lines (RLLs) and aseptic loosening. Hence, this RCT aimed to compare cemented Metasul MOM bearing with MoP and ceramic-on-polyethylene (CoP) bearings.

Qual era la domanda di ricerca principale?

How do long term outcomes, such as incidences of radiolucent lines (RLLs), rate of revision, and clinical results, compare between cemented Metasul MoM, MoP, and ceramic-on-polyethylene (CoP) bearings used in total hip replacement?

Caratteristiche dello studio +
Population:
374 (399 THRs; 341 hips at 7 year follow-up) patients, younger than 73 years of age, underwent THR for degenerative joint disease; 23 patients underwent bilateral procedures (none of which were simultaneous). All THRs were performed using the direct lateral approach with the patient in the lateral position using a triple-tapered and highly polished MS-30 femoral component (Sulzer Medica). Antibiotic-laden cement (Palacos with gentamicin; Heraeus, Hanau, Germany) was used for fixation.
Intervention:
MoM Group: A Weber polyethylene acetabular component with a 28 mm diameter Metasul insert (Sulzer Medica) was used with a Metasul modular head on patients (n=129 hips, 123 patients) (39M/84F; Mean age: 63.3 years (40.7 to 72.9))
Comparison:
MoP Group: A Weber polyethylene component was used that articulated with a 28 mm metal (Protasul CoCrMo alloy; Sulzer Medica) head (n=137 hips, 127 patients) (39M/88F; Mean age: 62.8 years (24.5 to 72.7)) CoP group: A Weber polyethylene component was used that articulated a ceramic (Sulox Alumina; Sulzer Medica) head (n=131 hips, 124 patients) (51M/73F, Mean age: 63.9 years (30.6 to 73.8))
Outcomes:
Outcomes were the Harris hip score, presence of radiolucent lines (RLLs), and aseptic loosening
Methods:
RCT:Single Center
Time:
7 years (follow up at 2, 5, and 7 years)
Quali erano i risultati importanti?
  • No difference was found in mean abduction angle (p=0.69) and HHS (p=0.104) at 7 years in 341 hips (86%) between the three bearing types.
  • MoM hips had higher incidences of peri-acetabular radiolucent lines of greater and equal to 2mm (p<0.001),
  • 12 revisions were performed, 8 in the MoM group, 3 in the MoP group and 1 in the CoP group. There was a significant difference in the number of revisions between the MoM and Mop/Cop groups (p<0.001).
  • No aseptic loosening was seen in the MoP and CoP groups.
  • 4infections occurred in the MoM group compared with 1 infection in the MoP group and 1 infection in the CoP group.
Che cosa devo ricordare di più?

The trial indicated that at a 7 year follow up period, cemented metal-on-metal bearings resulted in higher incidence of radioluscent lines around the acetabular component than metal-on-polyethylene and ceramic-on-polyethylene bearings. The metal-on-metal bearing also required higher revision rates and had lower Harris Hip scores.

Come influenzerà l'assistenza ai miei pazienti?

The study suggests that patients with 28 mm Metasul metal-on-metal bearing should be followed with caution when compared to the other implants used. It is unclear how these results may be applied to other metal on metal implants. Continuing surveillance of the outcomes of differing implant designs and bearing surfaces is warranted.

DISCLAIMER

Il contenuto di questa pagina è solo a scopo informativo e non intende sostituire la consulenza, la diagnosi o il trattamento medico professionale. Se ha bisogno di cure mediche, si rivolga sempre al suo medico o al pronto soccorso più vicino. Le opinioni, le convinzioni e i punti di vista espressi dalle persone sui contenuti presenti in questa pagina non riflettono le opinioni, le convinzioni e i punti di vista di OrthoEvidence.

0 di 4 articoli mensili GRATUITI sbloccati
Ha raggiunto il suo limite di 4 visualizzazioni di articoli gratuiti questo mese

Accesso a OrthoEvidence a soli 1,99 dollari a settimana.

Rimanga in contatto con le ultime prove. Può disdire in qualsiasi momento.
  • Valutazioni critiche dei più recenti studi randomizzati controllati ad alto impatto e delle revisioni sistematiche in ortopedia.
  • Accesso ai contenuti del podcast OrthoEvidence, tra cui collaborazioni con il Journal of Bone and Joint Surgery, interviste con chirurghi di fama internazionale e tavole rotonde su notizie e argomenti ortopedici.
  • Abbonamento a The Pulse, una newsletter bisettimanale basata sulle evidenze scientifiche, progettata per aiutarla a prendere decisioni cliniche migliori.
Upgrade
Close Dialog
Bentornato!
Ha dimenticato la password?
Inizi la sua prova gratuita oggi stesso!

Il suo account sarà affiliato a
e includerà l'accesso gratuito a OrthoEvidence


O
Ha dimenticato la password?

O
Controlli la sua e-mail

Se esiste un account con l'indirizzo e-mail fornito, le verrà inviata un'e-mail di reimpostazione della password. Se non vede l'e-mail, controlli la cartella spam o posta indesiderata.

Per ulteriore assistenza, contatti il nostro team di supporto.

Effettui il login per abilitare questa funzione

Per accedere a questa funzione, deve essere collegato ad un account OrthoEvidence attivo. Effettui il login o crei un account di prova GRATUITO.

Tradurre il Rapporto ACE

OrthoEvidence utilizza un servizio di traduzione di terze parti per rendere i contenuti accessibili in più lingue. Si prega di notare che, sebbene venga fatto ogni sforzo per garantire l'accuratezza, le traduzioni potrebbero non essere sempre perfette.

Come citare questo documento ACE Report

OrthoEvidence. Metasul metal-on-metal THRs have high rate of revision and incidences of radiolucent lines. OE Journal. 2013;1(16):1. Available from: https://myorthoevidence.com/AceReport/Show/metasul-metal-on-metal-thrs-have-high-rate-of-revision-and-incidences-of-radiolucent-lines

Copiare la citazione
Effettui il login per abilitare questa funzione

Per accedere a questa funzione, deve essere collegato ad un account OrthoEvidence attivo. Effettui il login o crei un account di prova GRATUITO.

Funzionalità per i membri Premium

Per accedere a questa funzione, deve essere collegato ad un account OrthoEvidence Premium.

Condividi questo articolo ACE Report