Better clinicoradiological results at 3 years with SpineJack vertebral augmentation vs BPK for VCFs .
Long-term safety and clinical performance of kyphoplasty and SpineJack(R) procedures in the treatment of osteoporotic vertebral compression fractures: a pilot, monocentric, investigator-initiated study
Osteoporos Int. 2019 Mar;30(3):637-645. doi: 10.1007/s00198-018-4773-530 patients with 1-2 osteoporotic vertebral compression fractures were randomized to vertebral augmentation surgery with either the SpineJack® implant or balloon kyphoplasty with the KyphX Xpander® implant. The purpose of the current study was to determine and compare 3-year outcome between groups. Clinical scores of visual analog pain scale and Oswestry Disability Index significantly favoured the SpineJack group compared to balloon kyphoplasty group. In addition, correction in anterior and middle vertebral body height ratio and kyphosis significantly favoured the SpineJack group. However, the comparisons for vertebral body height ratios may have been confounded by marked differences between groups at baseline.
افتح تقرير ACE الكامل
لديك إمكانية الوصول إلى المزيد من المقالات المجانية هذا الشهر، انقر أدناه لفتح هذا ACE Reports
افتح الآنتقييمات نقدية لأحدث التجارب المعشاة ذات الشواهد عالية التأثير والمراجعات المنهجية في جراحة العظام
الوصول إلى محتوى المدونة الصوتية OrthoEvidence، بما في ذلك التعاون مع مجلة جراحة العظام والمفاصل، ومقابلات مع جراحين معترف بهم دوليًا، ومناقشات المائدة المستديرة حول أخبار وموضوعات جراحة العظام
الاشتراك في النشرة الإخبارية The Pulse، وهي نشرة إخبارية مبنية على الأدلة مرتين أسبوعيًا مصممة لمساعدتك في اتخاذ قرارات سريرية أفضل
إمكانية الوصول الحصري إلى مقالات المحتوى الأصلي، بما في ذلك المراجعات المنهجية الداخلية، ومقالات عن طرق البحث الصحي وموضوعات جراحة العظام الساخنة